BACKGROUND: The quality and safety of advanced therapy products must be maintained throughout their production and quality control cycle to ensure their final use in patients. We validated the cell count method according to the International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use and European Pharmacopoeia, considering the tests' accuracy, precision, repeatability, linearity and range. METHODS: As the cell count is a potency test, we checked accuracy, precision, and linearity, according to ICH Q2. Briefly our experimental approach was first to evaluate the accuracy of Fast Read 102® compared to the Bürker chamber. Once the accuracy of the alternative method was demonstrated, we checked the precision and linearity test only using Fast Read 102®. The data were statistically analyzed by average, standard deviation and coefficient of variation percentages inter and intra operator. RESULTS: All the tests performed met the established acceptance criteria of a coefficient of variation of less than ten percent. For the cell count, the precision reached by each operator had a coefficient of variation of less than ten percent (total cells) and under five percent (viable cells). The best range of dilution, to obtain a slope line value very similar to 1, was between 1:8 and 1:128. CONCLUSIONS: Our data demonstrated that the Fast Read 102® count method is accurate, precise and ensures the linearity of the results obtained in a range of cell dilution. Under our standard method procedures, this assay may thus be considered a good quality control method for the cell count as a batch release quality control test. Moreover, the Fast Read 102® chamber is a plastic, disposable device that allows a number of samples to be counted in the same chamber. Last but not least, it overcomes the problem of chamber washing after use and so allows a cell count in a clean environment such as that in a Cell Factory. In a good manufacturing practice setting the disposable cell counting devices will allow a single use of the count chamber they can then be thrown away, thus avoiding the waste disposal of vital dye (e.g. Trypan Blue) or lysing solution (e.g. Tuerk solution).

Validation of analytical methods in GMP: The disposable Fast Read 102® device, an alternative practical approach for cell counting

GUNETTI, Monica;CASTIGLIA, Sara;MARESCHI, Katia;SANAVIO, Fiorella;MURARO, MICHELA;CASTELLO, LAURA MARISA;FERRERO, Ivana;FAGIOLI, FRANCA
2012-01-01

Abstract

BACKGROUND: The quality and safety of advanced therapy products must be maintained throughout their production and quality control cycle to ensure their final use in patients. We validated the cell count method according to the International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use and European Pharmacopoeia, considering the tests' accuracy, precision, repeatability, linearity and range. METHODS: As the cell count is a potency test, we checked accuracy, precision, and linearity, according to ICH Q2. Briefly our experimental approach was first to evaluate the accuracy of Fast Read 102® compared to the Bürker chamber. Once the accuracy of the alternative method was demonstrated, we checked the precision and linearity test only using Fast Read 102®. The data were statistically analyzed by average, standard deviation and coefficient of variation percentages inter and intra operator. RESULTS: All the tests performed met the established acceptance criteria of a coefficient of variation of less than ten percent. For the cell count, the precision reached by each operator had a coefficient of variation of less than ten percent (total cells) and under five percent (viable cells). The best range of dilution, to obtain a slope line value very similar to 1, was between 1:8 and 1:128. CONCLUSIONS: Our data demonstrated that the Fast Read 102® count method is accurate, precise and ensures the linearity of the results obtained in a range of cell dilution. Under our standard method procedures, this assay may thus be considered a good quality control method for the cell count as a batch release quality control test. Moreover, the Fast Read 102® chamber is a plastic, disposable device that allows a number of samples to be counted in the same chamber. Last but not least, it overcomes the problem of chamber washing after use and so allows a cell count in a clean environment such as that in a Cell Factory. In a good manufacturing practice setting the disposable cell counting devices will allow a single use of the count chamber they can then be thrown away, thus avoiding the waste disposal of vital dye (e.g. Trypan Blue) or lysing solution (e.g. Tuerk solution).
2012
10
1
112
123
http://www.translational-medicine.com/content/10/1/112
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3502295/
Cell count; Cell factory; Cell therapy; GMP; Validation methods; Humans; Reference Values; Reproducibility of Results; Laboratories; Quality Control; Biochemistry, Genetics and Molecular Biology (all); Medicine (all)
Gunetti, Monica; Castiglia, Sara; Rustichelli, Deborah; Mareschi, Katia; Sanavio, Fiorella; Muraro, Michela; Signorino, Elena; Castello, Laura; Ferrero, Ivana; Fagioli, Franca
File in questo prodotto:
File Dimensione Formato  
Gunetti_2012.pdf

Accesso aperto

Descrizione: Articolo principale
Tipo di file: PDF EDITORIALE
Dimensione 3.19 MB
Formato Adobe PDF
3.19 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1530739
Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 27
  • ???jsp.display-item.citation.isi??? 21
social impact