A single-crystal CVD (Chemical Vapor Deposition) diamond detector was used to measure γ rays in order to assess its performance in terms of energy resolution and linearity. For this purpose, 57Co, 133Ba, 22Na, 207Bi and 137Cs γ sources were used. Electrons scattered by the backward Compton process were detected in the diamond, in coincidence with (backscattered) γs measured in a NaI detector, placed at 180° from the CVD diamond detector with respect to the source. The resulting calibration shows a linear dependence of the charge deposited in the diamond and a resolution of about 24 keV FWHM for the energy of the incident γs between 40 keV (57Co) and 477 keV (137Cs), comparable with the resolution of our electronic chain.
Characterisation of scCVD diamond detectors with γ sources
AMAPANE, Nicola Carlo;ARGIRO', Stefano;BEOLE', Stefania;MASERA, Massimo;MILA, Giorgia;
2014-01-01
Abstract
A single-crystal CVD (Chemical Vapor Deposition) diamond detector was used to measure γ rays in order to assess its performance in terms of energy resolution and linearity. For this purpose, 57Co, 133Ba, 22Na, 207Bi and 137Cs γ sources were used. Electrons scattered by the backward Compton process were detected in the diamond, in coincidence with (backscattered) γs measured in a NaI detector, placed at 180° from the CVD diamond detector with respect to the source. The resulting calibration shows a linear dependence of the charge deposited in the diamond and a resolution of about 24 keV FWHM for the energy of the incident γs between 40 keV (57Co) and 477 keV (137Cs), comparable with the resolution of our electronic chain.File | Dimensione | Formato | |
---|---|---|---|
CVD_article-1.pdf
Open Access dal 02/08/2016
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
539.95 kB
Formato
Adobe PDF
|
539.95 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.