We aim at an analysis of the effects mechanical ventilators (MVs) and thoracic artificial lungs (TALs) will have on the cardiovascular system, especially on important quantities, such as left and right ventricular external work (EW), pressure-volume area (PVA) and cardiac mechanical efficiency (CME). Our analyses are based on simulation studies which were carried out by using our CARDIOSIM© software simulator. At first, we carried out simulation studies of patients undergoing mechanical ventilation (MV) without a thoracic artificial lung (TAL). Subsequently, we conducted simulation studies of patients who had been provided with a TAL, but did not undergo MV. We aimed at describing the patient's physiological characteristics and their variations with time, such as EW, PVA, CME, cardiac output (CO) and mean pulmonary arterial/venous pressure (PAP/PVP). We were starting with a simulation run under well-defined initial conditions which was followed by simulation runs for a wide range of mean intrathoracic pressure settings. Our simulations of MV without TAL showed that for mean intrathoracic pressure settings from negative (-4 mmHg) to positive (+5 mmHg) values, the left and right ventricular EW and PVA, right ventricular CME and CO decreased, whereas left ventricular CME and the PAP increased. The simulation studies of patients with a TAL, comprised all the usual TAL arrangements, viz. configurations “in series” and in parallel with the natural lung and, moreover, hybrid configurations. The main objective of the simulation studies was, as before, the assessment of the hemodynamic response to the application of a TAL. We could for instance show that, in case of an “in series” configuration, a reduction (an increase) in left (right) ventricular EW and PVA values occurred, whereas the best performance in terms of CO can be achieved in the case of an in parallel configuration.

Lung assist devices influence cardio-energetic parameters: Numerical simulation study

ALLASIA, Giampietro;DE ROSSI, Alessandra;CAVORETTO, Roberto;VENTURINO, Ezio;
2015-01-01

Abstract

We aim at an analysis of the effects mechanical ventilators (MVs) and thoracic artificial lungs (TALs) will have on the cardiovascular system, especially on important quantities, such as left and right ventricular external work (EW), pressure-volume area (PVA) and cardiac mechanical efficiency (CME). Our analyses are based on simulation studies which were carried out by using our CARDIOSIM© software simulator. At first, we carried out simulation studies of patients undergoing mechanical ventilation (MV) without a thoracic artificial lung (TAL). Subsequently, we conducted simulation studies of patients who had been provided with a TAL, but did not undergo MV. We aimed at describing the patient's physiological characteristics and their variations with time, such as EW, PVA, CME, cardiac output (CO) and mean pulmonary arterial/venous pressure (PAP/PVP). We were starting with a simulation run under well-defined initial conditions which was followed by simulation runs for a wide range of mean intrathoracic pressure settings. Our simulations of MV without TAL showed that for mean intrathoracic pressure settings from negative (-4 mmHg) to positive (+5 mmHg) values, the left and right ventricular EW and PVA, right ventricular CME and CO decreased, whereas left ventricular CME and the PAP increased. The simulation studies of patients with a TAL, comprised all the usual TAL arrangements, viz. configurations “in series” and in parallel with the natural lung and, moreover, hybrid configurations. The main objective of the simulation studies was, as before, the assessment of the hemodynamic response to the application of a TAL. We could for instance show that, in case of an “in series” configuration, a reduction (an increase) in left (right) ventricular EW and PVA values occurred, whereas the best performance in terms of CO can be achieved in the case of an in parallel configuration.
2015
Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
Institute of Electrical and Electronics Engineers Inc.
2015
4515
4519
9781424492718
9781424492718
Signal Processing; Biomedical Engineering; Health Informatics
De Lazzari, C.; Quatember, B.; Recheis, W.; Mayr, M.; Demertzis, S.; Allasia, G.; De Rossi, A.; Cavoretto, R.; Venturino, E.; Genuini, I.
File in questo prodotto:
File Dimensione Formato  
P22.pdf

Accesso aperto

Tipo di file: PDF EDITORIALE
Dimensione 1.07 MB
Formato Adobe PDF
1.07 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1597599
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact