PURPOSE: Left atrial volume (LA) and pulmonary vein (PV) anatomy may potentially relate to technical challenges in achieving stable and effective catheter position in case of atrial fibrillation (AF) ablation by means of "one-shot" catheters. The aim of this study was to investigate whether LA volume and PV anatomy, evaluated by computed tomography (CT) or magnetic resonance (MR) prior to ablation, predict acute and midterm outcome of AF ablation by nMARQ™. METHODS: We included 75 patients (mean age 58 ± 11 years, 67 % male) with symptomatic paroxysmal AF. All patients underwent CT/MR scanning prior to catheter ablation to evaluate LA volume and PV anatomy. All the patients underwent PV isolation by nMARQ™, an open-irrigated mapping and radiofrequency (RF) decapolar ablation catheter. Ablation was guided by electroanatomic mapping allowing RF energy delivery in the antral region of PVs from ten irrigated electrodes simultaneously. RESULTS: Mean LA volume was 75 ± 40 ml. A normal anatomy (4 PVs) was documented in 40 (53 %) patients and abnormal anatomy (common truncus or accessory PVs) in 35 patients. Mean procedural and fluoroscopy times were 94 ± 55 and 8 ± 5 min, respectively, without significant differences among patients with normal or abnormal anatomy (92 ± 45 vs 95 ± 64 min, p = 0.85 and 6 ± 3 vs 8 ± 4 min, p = 0.65, respectively). Mean ablation time was 14 ± 3 min, and 99 % of the targeted veins were isolated with a mean of 23 ± 5 RF pulses per patient. After a mean follow-up of 17 ± 8 months, 23 (31 %) patients had an atrial arrhythmia recurrence. Neither LA volume nor PV anatomy was a predictor of outcome. CONCLUSIONS: LA volume and PV anatomy did not affect procedural data and outcome in patients who underwent PV isolation by an open-irrigated mapping and RF decapolar ablation catheter.

Effect of left atrial volume and pulmonary vein anatomy on outcome of nMARQ™ catheter ablation of paroxysmal atrial fibrillation

ANSELMINO, Matteo;GAITA, Fiorenzo
2017-01-01

Abstract

PURPOSE: Left atrial volume (LA) and pulmonary vein (PV) anatomy may potentially relate to technical challenges in achieving stable and effective catheter position in case of atrial fibrillation (AF) ablation by means of "one-shot" catheters. The aim of this study was to investigate whether LA volume and PV anatomy, evaluated by computed tomography (CT) or magnetic resonance (MR) prior to ablation, predict acute and midterm outcome of AF ablation by nMARQ™. METHODS: We included 75 patients (mean age 58 ± 11 years, 67 % male) with symptomatic paroxysmal AF. All patients underwent CT/MR scanning prior to catheter ablation to evaluate LA volume and PV anatomy. All the patients underwent PV isolation by nMARQ™, an open-irrigated mapping and radiofrequency (RF) decapolar ablation catheter. Ablation was guided by electroanatomic mapping allowing RF energy delivery in the antral region of PVs from ten irrigated electrodes simultaneously. RESULTS: Mean LA volume was 75 ± 40 ml. A normal anatomy (4 PVs) was documented in 40 (53 %) patients and abnormal anatomy (common truncus or accessory PVs) in 35 patients. Mean procedural and fluoroscopy times were 94 ± 55 and 8 ± 5 min, respectively, without significant differences among patients with normal or abnormal anatomy (92 ± 45 vs 95 ± 64 min, p = 0.85 and 6 ± 3 vs 8 ± 4 min, p = 0.65, respectively). Mean ablation time was 14 ± 3 min, and 99 % of the targeted veins were isolated with a mean of 23 ± 5 RF pulses per patient. After a mean follow-up of 17 ± 8 months, 23 (31 %) patients had an atrial arrhythmia recurrence. Neither LA volume nor PV anatomy was a predictor of outcome. CONCLUSIONS: LA volume and PV anatomy did not affect procedural data and outcome in patients who underwent PV isolation by an open-irrigated mapping and RF decapolar ablation catheter.
2017
48
2
201
207
www.wkap.nl/journalhome.htm/1383-875X
Anatomy; Atrial fibrillation; Catheter ablation; Midterm outcome; Multielectrode ablation catheter; Safety; Cardiology and Cardiovascular Medicine; Physiology (medical)
Stabile, Giuseppe; Anselmino, Matteo; Soldati, Ezio; de Ruvo, Ermengildo; Solimene, Francesco; Iuliano, Assunta; Sciarra, Luigi; Bongiorni, Maria Graz...espandi
File in questo prodotto:
File Dimensione Formato  
Final accepted JICE with figures_4aperto.pdf

Accesso aperto

Tipo di file: POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione 1.68 MB
Formato Adobe PDF
1.68 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1615588
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
social impact