An optrode for continuous monitoring of potassium to be used in whole blood was realized. Several commercial chromoionophores have been tried and composition of the membrane has been optimized in order to obtain the highest sensitivity around pH 7.4 and maximum light stability. The sensor is sensitive to pH but a simple equation relates absorbance and the value (pH + log K) and it was used in real experiments with cow whole blood. The analytical range is 1–10 mM of K+ and it matches the clinical requirements.

An optimized optrode for continuous potassium monitoring in whole blood

CAPUTO, Giuseppe;
1999-01-01

Abstract

An optrode for continuous monitoring of potassium to be used in whole blood was realized. Several commercial chromoionophores have been tried and composition of the membrane has been optimized in order to obtain the highest sensitivity around pH 7.4 and maximum light stability. The sensor is sensitive to pH but a simple equation relates absorbance and the value (pH + log K) and it was used in real experiments with cow whole blood. The analytical range is 1–10 mM of K+ and it matches the clinical requirements.
1999
401
1-2
129
136
http://www.sciencedirect.com/science/article/pii/S0003267099005085
Chromoionophore; Continuous monitoring; Optrode; Potassium monitoring; Whole blood; Analytical Chemistry; Environmental Chemistry; Biochemistry; Spectroscopy
Malavolti, E.; Cagnini, A.; Caputo, G.; Della Ciana, L.; Mascini, M
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1632315
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 11
  • ???jsp.display-item.citation.isi??? ND
social impact