This method describes a sensitive, specific, reliable and reproducible reverse phase high-performance liquid chromatography (RP-HPLC) assay developed and validated for the quantification of extracellular purine nucleotides and nucleosides produced by purified chronic lymphocytic leukemia (CLL) cells under different culture conditions. The chromatographic separation of adenosine 5'-monophosphate (AMP), adenosine (ADO) and inosine (INO) is performed at RT on a silica-based, reversed-phase column that is used for polar compound retention. The method includes a binary mobile phase, which consists of 7 mM ammonium acetate and acetonitrile with a flow rate of 1.00 ml/min. The eluates are monitored using a Photodiode Array UV detector set at 260 nm. A standard calibration curve is generated to calculate the equation for the analytical quantification of each purine compound. System control, data acquisition and analysis are then performed. Applying this protocol, AMP, INO and ADO elute at 7, 11 and 11.9 min, respectively, and the total run time for each sample is 20 min. This protocol may be applied to different cell types and cell lines (both suspension and adherent), using culture media as matrix. The advantages are easy and fast sample preparation and the requirement of a small amount of supernatant for analysis. Furthermore, the use of a serum-free medium allows skipping the protein precipitation step with acetonitrile that impacts the final concentration of purine compounds. One of the limitations of the method is the requirement of the equilibration column run before each single sample run, making the total run time of the experiment longer and preventing high throughput screening applications.
HPLC-based assay to monitor extracellular nucleotide/nucleoside metabolism in human chronic lymphocytic leukemia cells
Serra, Sara;Deaglio, Silvia
Last
2016-01-01
Abstract
This method describes a sensitive, specific, reliable and reproducible reverse phase high-performance liquid chromatography (RP-HPLC) assay developed and validated for the quantification of extracellular purine nucleotides and nucleosides produced by purified chronic lymphocytic leukemia (CLL) cells under different culture conditions. The chromatographic separation of adenosine 5'-monophosphate (AMP), adenosine (ADO) and inosine (INO) is performed at RT on a silica-based, reversed-phase column that is used for polar compound retention. The method includes a binary mobile phase, which consists of 7 mM ammonium acetate and acetonitrile with a flow rate of 1.00 ml/min. The eluates are monitored using a Photodiode Array UV detector set at 260 nm. A standard calibration curve is generated to calculate the equation for the analytical quantification of each purine compound. System control, data acquisition and analysis are then performed. Applying this protocol, AMP, INO and ADO elute at 7, 11 and 11.9 min, respectively, and the total run time for each sample is 20 min. This protocol may be applied to different cell types and cell lines (both suspension and adherent), using culture media as matrix. The advantages are easy and fast sample preparation and the requirement of a small amount of supernatant for analysis. Furthermore, the use of a serum-free medium allows skipping the protein precipitation step with acetonitrile that impacts the final concentration of purine compounds. One of the limitations of the method is the requirement of the equilibration column run before each single sample run, making the total run time of the experiment longer and preventing high throughput screening applications.File | Dimensione | Formato | |
---|---|---|---|
JoVE accepted manuscript.pdf
Accesso riservato
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
146.01 kB
Formato
Adobe PDF
|
146.01 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.