Background: To evaluate the early safety and clinical performance of the new low-profile RelayPro Thoracic Stent-Graft System in patients with thoracic aortic disease. Methods: This was an international, prospective, single-arm study in patients diagnosed with thoracic aorta disease (aneurysm, pseudoaneurysm, dissection, penetrating atherosclerotic ulcer, or intramural hematoma) and treated with a RelayPro stent-graft (in bare stent and/or nonbare stent configurations). The primary endpoints were freedom from aneurysm or dissection-related mortality and stent-graft performance. Results: A total of 31 patients were treated with the RelayPro thoracic stent-graft between 2014 and 2015 at 8 sites in Italy and Spain. Mean age was 72.1 (±10.2) years and 77% were male, 74% with hypertension, and 42% with a history of smoking. Twenty-four (77%) had aneurysms (fusiform in 46%, saccular in 42%, pseudoaneurysm in 12%); 5 (16%) had penetrating atherosclerotic ulcer; and 2 (6%) had chronic Type B dissection. Mean vascular access diameter was 9.1 mm (6–13 mm); 7 patients (23%) had vascular access of 7 mm or less. Technical success was 100% (primary, 90%; assisted primary, 10%). Freedom from aneurysm/dissection-related mortality through 30 days was 100%. Freedom from device-related major adverse events through 30 days was 94%. At 1 year, there was 1 (3%) type Ib and 1 (3%) type II endoleak, 1 (3%) nonaneurysm-related late death, and 1 (3%) secondary intervention (to correct type Ib endoleak). Conclusions: The RelayPro has a 3–4 French profile reduction to allow endovascular repair of thoracic aortic disease in patients with smaller anatomies. This study shows good initial stent-graft performance and a favorable early safety profile.
Prospective multicenter study of the low-profile Relay stent-graft in patients with thoracic aortic disease: the RE-GENERATION study
Verzini, Fabio
Last
2019-01-01
Abstract
Background: To evaluate the early safety and clinical performance of the new low-profile RelayPro Thoracic Stent-Graft System in patients with thoracic aortic disease. Methods: This was an international, prospective, single-arm study in patients diagnosed with thoracic aorta disease (aneurysm, pseudoaneurysm, dissection, penetrating atherosclerotic ulcer, or intramural hematoma) and treated with a RelayPro stent-graft (in bare stent and/or nonbare stent configurations). The primary endpoints were freedom from aneurysm or dissection-related mortality and stent-graft performance. Results: A total of 31 patients were treated with the RelayPro thoracic stent-graft between 2014 and 2015 at 8 sites in Italy and Spain. Mean age was 72.1 (±10.2) years and 77% were male, 74% with hypertension, and 42% with a history of smoking. Twenty-four (77%) had aneurysms (fusiform in 46%, saccular in 42%, pseudoaneurysm in 12%); 5 (16%) had penetrating atherosclerotic ulcer; and 2 (6%) had chronic Type B dissection. Mean vascular access diameter was 9.1 mm (6–13 mm); 7 patients (23%) had vascular access of 7 mm or less. Technical success was 100% (primary, 90%; assisted primary, 10%). Freedom from aneurysm/dissection-related mortality through 30 days was 100%. Freedom from device-related major adverse events through 30 days was 94%. At 1 year, there was 1 (3%) type Ib and 1 (3%) type II endoleak, 1 (3%) nonaneurysm-related late death, and 1 (3%) secondary intervention (to correct type Ib endoleak). Conclusions: The RelayPro has a 3–4 French profile reduction to allow endovascular repair of thoracic aortic disease in patients with smaller anatomies. This study shows good initial stent-graft performance and a favorable early safety profile.File | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S0890509619300342-main.pdf
Accesso riservato
Descrizione: post print articolo
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
1.68 MB
Formato
Adobe PDF
|
1.68 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.