This is a report of a prospective study to evaluate the safety and efficacy of the Medtronic AneuRx(TM) stent-graft. Patients with an infrarenal aneurysm with a proximal neck length of greater than 10 mm and a neck diameter not greater than 26 mm and iliac artery diameters of at least 6 mm were accepted for endovascular repair using the Medtronic AneuRx(RM) modular stent-graft, A total of 104 patients were included in the study. Transfemoral placement of the bifurcated stent-graft was successful in 102 of 104 patients. Two conversions were performed. The mean operating time was 148 min (range 75-480) and the mean blood loss was 605 mi (range 100-2900). The mean follow-up was 15 months (range 12-21). Complications were rare and no kinking or migration occurred. At follow-up, four endoleaks persisted after 12 months, The Medtronic AneuRx(TM) stent-graft is a safe and efficacious alternative to open abdominal aortic aneurysm repair with excellent early results. (C) 1999 The International Society for Cardiovascular Surgery, Published by Elsevier Science Ltd. All rights reserved.

Endoluminal repair of infrarenal abdominal aortic aneurysms using a modular stent-graft: one year clinical results from a european multicentre trial

Verzini F;
1999-01-01

Abstract

This is a report of a prospective study to evaluate the safety and efficacy of the Medtronic AneuRx(TM) stent-graft. Patients with an infrarenal aneurysm with a proximal neck length of greater than 10 mm and a neck diameter not greater than 26 mm and iliac artery diameters of at least 6 mm were accepted for endovascular repair using the Medtronic AneuRx(RM) modular stent-graft, A total of 104 patients were included in the study. Transfemoral placement of the bifurcated stent-graft was successful in 102 of 104 patients. Two conversions were performed. The mean operating time was 148 min (range 75-480) and the mean blood loss was 605 mi (range 100-2900). The mean follow-up was 15 months (range 12-21). Complications were rare and no kinking or migration occurred. At follow-up, four endoleaks persisted after 12 months, The Medtronic AneuRx(TM) stent-graft is a safe and efficacious alternative to open abdominal aortic aneurysm repair with excellent early results. (C) 1999 The International Society for Cardiovascular Surgery, Published by Elsevier Science Ltd. All rights reserved.
1999
7
5
503
507
https://vpn.unipg.it/science/article/pii/,DanaInfo=www.sciencedirect.com+S0967210999000186
Cardiac & Cardiovascular Systems; Surgery; abdominal aortic aneurysms; endovascular repair
Tutein Nolthenius Rp; van der Berg Jc; Biasi Gm; Piglionica Mr; Meregaglia D; Ferrari Sa; Coppi G; Pacchioni R; Gennai S; Cao P; Barzi F; Verzini F; Maselli A; Caporali S; Moll Fl
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1693237
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 12
social impact