Intermittent Pneumatic Compression (IPC) devices can be used to analyze the mechanisms underlying several vascular phenomena, such as hyperaemia. Commercial devices have limited dynamics and do not allow the delivery of customizable compressive pressure patterns, making the analysis of such phenomena difficult, which may require the application of long stimulations with low amplitude as well as fast compressions with higher pressure level. To overcome these issues, a novel pneumotronic device aimed to the investigation of the physiological effects induced by limb compressions is conceived and presented in this work. The design requirements of the system, capable of delivering customizable compressive patterns in the range 0 mmHg - 200 mmHg, are outlined. The final prototype architecture is described, and a mathematical model of the entire system, also including the interaction between the device and the limb tissues, is proposed. The performance of the device was evaluated in several conditions by means of simulations, whose results were compared to the data collected from experimental trials in order to validate the model. The outcomes of both experimentation and simulation trials proved the effectiveness of the solution proposed. A possible employment of this device for the investigation of the rapid compression-induced hyperaemia is presented. Other potential applications concern the wide range of intermittent-pneumatic compression treatments.

Design and Simulation of a Novel Pneumotronic System Aimed to the Investigation of Vascular Phenomena Induced by Limb Compression

Messere A.;Pertusio R.;Roatta S.
Last
2019-01-01

Abstract

Intermittent Pneumatic Compression (IPC) devices can be used to analyze the mechanisms underlying several vascular phenomena, such as hyperaemia. Commercial devices have limited dynamics and do not allow the delivery of customizable compressive pressure patterns, making the analysis of such phenomena difficult, which may require the application of long stimulations with low amplitude as well as fast compressions with higher pressure level. To overcome these issues, a novel pneumotronic device aimed to the investigation of the physiological effects induced by limb compressions is conceived and presented in this work. The design requirements of the system, capable of delivering customizable compressive patterns in the range 0 mmHg - 200 mmHg, are outlined. The final prototype architecture is described, and a mathematical model of the entire system, also including the interaction between the device and the limb tissues, is proposed. The performance of the device was evaluated in several conditions by means of simulations, whose results were compared to the data collected from experimental trials in order to validate the model. The outcomes of both experimentation and simulation trials proved the effectiveness of the solution proposed. A possible employment of this device for the investigation of the rapid compression-induced hyperaemia is presented. Other potential applications concern the wide range of intermittent-pneumatic compression treatments.
2019
16
3
550
562
https://rd.springer.com/journal/42235
human-machine interaction; hyperaemia; limb compression; pneumatic system modeling; pneumotronic system; pneutronics
Ferraresi C.; De Benedictis C.; Maffiodo D.; Franco W.; Messere A.; Pertusio R.; Roatta S.
File in questo prodotto:
File Dimensione Formato  
2019 Ferraresi et al, J Bionic Eng, Design and simulation of pneumotronic system.pdf

Accesso riservato

Descrizione: articolo principale
Tipo di file: PDF EDITORIALE
Dimensione 863.79 kB
Formato Adobe PDF
863.79 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
2019 Ferraresi et al JBE postprint.pdf

Open Access dal 23/05/2021

Descrizione: postprint articolo principale
Tipo di file: POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione 2.47 MB
Formato Adobe PDF
2.47 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1727356
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 9
social impact