Purpose: Ultrasound (US) is a useful tool for peripheral vein cannulation in patients with difficult venous access. However, few data about the survival of US-guided peripheral catheters in acute care setting exist. Some studies showed that the survival rate of standard-length catheters (SC) is poor especially in obese patients. The use of longer than normal catheters could provide a solution to low survival rate. The aim of the present study was to compare US-guided peripheral SCs vs US-guided peripheral long catheters inserted with Seldinger technique (LC) in acute hospitalized patients with difficult venous access. Methods: This was a prospective, randomized controlled trial. A total of 100 consecutively admitted subjects in an urban High Dependency Unit were randomized to obtain US-guided intravenous access using either SC or LC after 3 failed blind attempts. Primary outcome was catheter failure rate. Results: Success rate was 86% in the SC groups and 84% in the LC group (P = .77). Time requested to positioning venous access resulted to be shorter for SC as opposed to LC (9.5 vs 16.8 minutes, respectively; P = .001). Catheter failure was observed in 45% of patients in the SC group and in 14% of patients in the LC group (relative risk, 3.2; P < .001). Conclusions: Both SC and LC US-guided cannulations have a high success rate in patients with difficult venous access. Notwithstanding a higher time to cannulation, LC US-guided procedure is associated with a lower risk of catheter failure compared with SC US-guided procedure. © 2012 Elsevier Inc. All rights reserved.

Standard-length catheters vs long catheters in ultrasound-guided peripheral vein cannulation

Elia F.;De Filippi G.;Milan A.;Apra F.
2012-01-01

Abstract

Purpose: Ultrasound (US) is a useful tool for peripheral vein cannulation in patients with difficult venous access. However, few data about the survival of US-guided peripheral catheters in acute care setting exist. Some studies showed that the survival rate of standard-length catheters (SC) is poor especially in obese patients. The use of longer than normal catheters could provide a solution to low survival rate. The aim of the present study was to compare US-guided peripheral SCs vs US-guided peripheral long catheters inserted with Seldinger technique (LC) in acute hospitalized patients with difficult venous access. Methods: This was a prospective, randomized controlled trial. A total of 100 consecutively admitted subjects in an urban High Dependency Unit were randomized to obtain US-guided intravenous access using either SC or LC after 3 failed blind attempts. Primary outcome was catheter failure rate. Results: Success rate was 86% in the SC groups and 84% in the LC group (P = .77). Time requested to positioning venous access resulted to be shorter for SC as opposed to LC (9.5 vs 16.8 minutes, respectively; P = .001). Catheter failure was observed in 45% of patients in the SC group and in 14% of patients in the LC group (relative risk, 3.2; P < .001). Conclusions: Both SC and LC US-guided cannulations have a high success rate in patients with difficult venous access. Notwithstanding a higher time to cannulation, LC US-guided procedure is associated with a lower risk of catheter failure compared with SC US-guided procedure. © 2012 Elsevier Inc. All rights reserved.
2012
30
5
712
716
Aged; Catheterization, Peripheral; Critical Care; Equipment Failure; Female; Humans; Male; Time Factors; Ultrasonography, Interventional; Catheters, Indwelling
Elia F.; Ferrari G.; Molino P.; Converso M.; De Filippi G.; Milan A.; Apra F.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1728411
Citazioni
  • ???jsp.display-item.citation.pmc??? 12
  • Scopus 81
  • ???jsp.display-item.citation.isi??? 69
social impact