Due to be launched in late 2021, the Imaging X-ray Polarimetry Explorer (IXPE) is a NASA Small Explorer mission designed to perform polarization measurements in the 2–8 keV band, complemented with imaging, spectroscopy and timing capabilities. At the heart of the focal plane is a set of three polarization-sensitive Gas Pixel Detectors (GPD), each based on a custom ASIC acting as a charge-collecting anode. In this paper we shall review the design, manufacturing, and test of the IXPE focal-plane detectors, with particular emphasis on the connection between the science drivers, the performance metrics and the operational aspects. We shall present a thorough characterization of the GPDs in terms of effective noise, trigger efficiency, dead time, uniformity of response, and spectral and polarimetric performance. In addition, we shall discuss in detail a number of instrumental effects that are relevant for high-level science analysis—particularly as far as the response to unpolarized radiation and the stability in time are concerned.

Design, construction, and test of the Gas Pixel Detectors for the IXPE mission

Bonino R.;Maldera S.;Oppedisano C.;Orsini L.;
2021-01-01

Abstract

Due to be launched in late 2021, the Imaging X-ray Polarimetry Explorer (IXPE) is a NASA Small Explorer mission designed to perform polarization measurements in the 2–8 keV band, complemented with imaging, spectroscopy and timing capabilities. At the heart of the focal plane is a set of three polarization-sensitive Gas Pixel Detectors (GPD), each based on a custom ASIC acting as a charge-collecting anode. In this paper we shall review the design, manufacturing, and test of the IXPE focal-plane detectors, with particular emphasis on the connection between the science drivers, the performance metrics and the operational aspects. We shall present a thorough characterization of the GPDs in terms of effective noise, trigger efficiency, dead time, uniformity of response, and spectral and polarimetric performance. In addition, we shall discuss in detail a number of instrumental effects that are relevant for high-level science analysis—particularly as far as the response to unpolarized radiation and the stability in time are concerned.
2021
133
102628
1
16
https://arxiv.org/abs/2107.05496
Gas detectors; X-ray polarimetry
Baldini L.; Barbanera M.; Bellazzini R.; Bonino R.; Borotto F.; Brez A.; Caporale C.; Cardelli C.; Castellano S.; Ceccanti M.; Citraro S.; Di Lalla N....espandi
File in questo prodotto:
File Dimensione Formato  
Design-construction-test_GPD_ApP2021.pdf

Accesso aperto

Tipo di file: PDF EDITORIALE
Dimensione 3.3 MB
Formato Adobe PDF
3.3 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1805229
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 81
  • ???jsp.display-item.citation.isi??? 65
social impact