Objectives: The aim of this study was to compare the Electronic von Frey Anaesthesiometer (EVF) and the Small Animal ALGOmeter (SMALGO), used to measure sensory thresholds in 13 healthy cats at both the stifle and the lumbosacral joint, in terms of inter-rater and inter-device reliability. Methods: Two independent observers carried out the sets of measurements in a randomised order, with a 45 min interval between them, in each cat. The inter-rater and inter-device reliability were evaluated by calculating the inter-rater correlation coefficient (ICC) for each pair of measurements. The Bland–Altman method was used as an additional tool to assess the level of agreement between the two algometers. Results: The mean ± SD sensory thresholds measured with the EVF were 311 ± 116 g and 378 ± 178 g for the stifle and for the lumbosacral junction, respectively, whereas those measured with the SMALGO were 391 ±172 g and 476 ± 172 g. The inter-rater reliability was fair (ICC >0.4) for each pair of measurements except those taken at the level of the stifle with the SMALGO, for which the level of agreement between observers A and B was poor (ICC = 0.01). The inter-device reliability was good (ICC = 0.73; P = 0.001). The repetition of the measurements affected reliability, as the thresholds obtained after the 45 min break were consistently lower than those measured during the first part of the trial (P = 0.02). Conclusions and relevance: The EVF and the SMALGO may be used interchangeably in cats, especially when the area to be tested is the lumbosacral joint. However, when the thresholds are measured at the stifle, the inter-observer reliability is better with the EVF than with the SMALGO. The reliability decreases when the measurements are repeated within a short time interval, suggesting a limited clinical applicability of quantitative sensory testing with both algometers in cats.

Inter-rater and inter-device reliability of mechanical thresholds measurement with the Electronic von Frey Anaesthesiometer and the SMALGO in healthy cats

Adami C.;Lardone E.;
2019-01-01

Abstract

Objectives: The aim of this study was to compare the Electronic von Frey Anaesthesiometer (EVF) and the Small Animal ALGOmeter (SMALGO), used to measure sensory thresholds in 13 healthy cats at both the stifle and the lumbosacral joint, in terms of inter-rater and inter-device reliability. Methods: Two independent observers carried out the sets of measurements in a randomised order, with a 45 min interval between them, in each cat. The inter-rater and inter-device reliability were evaluated by calculating the inter-rater correlation coefficient (ICC) for each pair of measurements. The Bland–Altman method was used as an additional tool to assess the level of agreement between the two algometers. Results: The mean ± SD sensory thresholds measured with the EVF were 311 ± 116 g and 378 ± 178 g for the stifle and for the lumbosacral junction, respectively, whereas those measured with the SMALGO were 391 ±172 g and 476 ± 172 g. The inter-rater reliability was fair (ICC >0.4) for each pair of measurements except those taken at the level of the stifle with the SMALGO, for which the level of agreement between observers A and B was poor (ICC = 0.01). The inter-device reliability was good (ICC = 0.73; P = 0.001). The repetition of the measurements affected reliability, as the thresholds obtained after the 45 min break were consistently lower than those measured during the first part of the trial (P = 0.02). Conclusions and relevance: The EVF and the SMALGO may be used interchangeably in cats, especially when the area to be tested is the lumbosacral joint. However, when the thresholds are measured at the stifle, the inter-observer reliability is better with the EVF than with the SMALGO. The reliability decreases when the measurements are repeated within a short time interval, suggesting a limited clinical applicability of quantitative sensory testing with both algometers in cats.
2019
21
10
979
984
Chronic pain; mechanical thresholds; quantitative sensory testing; Animals; Cat Diseases; Cats; Female; Male; Neurologic Examination; Observer Variation; Reproducibility of Results; Sensory Thresholds
Adami C.; Lardone E.; Monticelli P.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1832922
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 5
social impact