Reliable post-approval surveillance of implantable cardioverter-defibrillator (ICD) lead performance remains a challenge. In the past, two ICD leads were recalled due to a high frequency of failures. In this meta-analysis, we sought to provide a combined estimate of failure-free rate for ICD leads by reconstructing individual patient data from published Kaplan-Meier (KM) curves and to investigate whether estimates could be influenced by the characteristics of the study.

Implantable cardioverter defibrillator lead performance: A systematic review and individual patient data Meta-analysis

Comoretto, Rosanna Irene;Gregori, Dario
2023-01-01

Abstract

Reliable post-approval surveillance of implantable cardioverter-defibrillator (ICD) lead performance remains a challenge. In the past, two ICD leads were recalled due to a high frequency of failures. In this meta-analysis, we sought to provide a combined estimate of failure-free rate for ICD leads by reconstructing individual patient data from published Kaplan-Meier (KM) curves and to investigate whether estimates could be influenced by the characteristics of the study.
2023
373
57
63
Failure; ICD leads; Implantable cardioverter-defibrillator; Post market surveillance; meta-analysis
Giacopelli, Daniele; Azzolina, Danila; Comoretto, Rosanna Irene; Quartieri, Fabio; Rovaris, Giovanni; Schillaci, Vincenzo; Gargaro, Alessio; Gregori, ...espandi
File in questo prodotto:
File Dimensione Formato  
Meta-analysis_ICD_lead.pdf

Accesso riservato

Descrizione: Manuscript
Tipo di file: PDF EDITORIALE
Dimensione 4.74 MB
Formato Adobe PDF
4.74 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1881944
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 2
social impact