The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Cyclosporine (CsA) immunoassay in 7 clinical laboratories in comparison to liquid chromatography/tandem mass spectrometry (LC/MS/MS), Abbott TDx, Cobas Integra 800, and the Dade Dimension Xpand immunoassay. The ARCHITECT assay uses a whole blood specimen, a pretreatment step with organic reagents to precipitate proteins and extract the drug, followed by a 2-step automated immunoassay with magnetic microparticles coated with anti-CsA antibody and an acridinium-CsA tracer. Imprecision testing at the 7 evaluation sites gave a range of total % coefficient of variations of 7.5%-12.2% at 87.5 ng/mL, 6.6%-14.3% at 411 ng/mL, and 5.2%-10.7% at 916 ng/mL. The lower limit of quantification ranged from 12 to 20 ng/mL. Purified CsA metabolites AM1, AM1c, AM4N, AM9, and AM19 were tested in whole blood by the ARCHITECT assay and showed minimal cross-reactivity at all 7 sites. In particular, AM1 and AM9 cross-reactivity in the ARCHITECT assay, ranged from-2.5% to 0.2% and-0.8% to 2.2%, respectively, and was significantly lower than for the TDx assay, in which the values were 3.2% and 16.1%, respectively. Comparable testing of metabolites in the Dade Dimension Xpand assay at 2 evaluation sites showed cross-reactivity to AM4N (6.4% and 6.8%) and AM9 (2.6% and 3.6%) and testing on the Roche Integra 800 showed cross-reactivity to AM1c (2.4%), AM9 (10.7%), and AM19 (2.8%). Cyclosporine International Proficiency Testing Scheme samples, consisting of both pooled specimens from patients receiving CsA therapy as well as whole-blood specimens supplemented with CsA, were tested by the ARCHITECT assay at 6 sites and showed an average bias of-24 to-58 ng/mL versus LC/MSMS CsA and-2 to-37 ng/mL versus AxSYM CsA. Studies were performed with the ARCHITECT CsA assay on patient specimens with the following results: ARCHITECT CsA assay versus LC/MSMS, average bias of 31 ng/mL; ARCHITECT versus the Dade Dimension assay (4 sites), average biases of-7 to-228 ng/mL; ARCHITECT versus AxSYM and TDx, average biases of-4 and-53 ng/mL, respectively. Spearman correlation coefficients were ≥0.89. The ARCHITECT CsA assay has significantly reduced CsA metabolite interference relative to other immunoassays and is a convenient and sensitive semiautomated method to measure CsA in whole blood. Copyright © 2010 by Lippincott Williams & Wilkins.

Multisite analytical evaluation of the abbott ARCHITECT cyclosporine assay

Mengozzi G.;
2010-01-01

Abstract

The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Cyclosporine (CsA) immunoassay in 7 clinical laboratories in comparison to liquid chromatography/tandem mass spectrometry (LC/MS/MS), Abbott TDx, Cobas Integra 800, and the Dade Dimension Xpand immunoassay. The ARCHITECT assay uses a whole blood specimen, a pretreatment step with organic reagents to precipitate proteins and extract the drug, followed by a 2-step automated immunoassay with magnetic microparticles coated with anti-CsA antibody and an acridinium-CsA tracer. Imprecision testing at the 7 evaluation sites gave a range of total % coefficient of variations of 7.5%-12.2% at 87.5 ng/mL, 6.6%-14.3% at 411 ng/mL, and 5.2%-10.7% at 916 ng/mL. The lower limit of quantification ranged from 12 to 20 ng/mL. Purified CsA metabolites AM1, AM1c, AM4N, AM9, and AM19 were tested in whole blood by the ARCHITECT assay and showed minimal cross-reactivity at all 7 sites. In particular, AM1 and AM9 cross-reactivity in the ARCHITECT assay, ranged from-2.5% to 0.2% and-0.8% to 2.2%, respectively, and was significantly lower than for the TDx assay, in which the values were 3.2% and 16.1%, respectively. Comparable testing of metabolites in the Dade Dimension Xpand assay at 2 evaluation sites showed cross-reactivity to AM4N (6.4% and 6.8%) and AM9 (2.6% and 3.6%) and testing on the Roche Integra 800 showed cross-reactivity to AM1c (2.4%), AM9 (10.7%), and AM19 (2.8%). Cyclosporine International Proficiency Testing Scheme samples, consisting of both pooled specimens from patients receiving CsA therapy as well as whole-blood specimens supplemented with CsA, were tested by the ARCHITECT assay at 6 sites and showed an average bias of-24 to-58 ng/mL versus LC/MSMS CsA and-2 to-37 ng/mL versus AxSYM CsA. Studies were performed with the ARCHITECT CsA assay on patient specimens with the following results: ARCHITECT CsA assay versus LC/MSMS, average bias of 31 ng/mL; ARCHITECT versus the Dade Dimension assay (4 sites), average biases of-7 to-228 ng/mL; ARCHITECT versus AxSYM and TDx, average biases of-4 and-53 ng/mL, respectively. Spearman correlation coefficients were ≥0.89. The ARCHITECT CsA assay has significantly reduced CsA metabolite interference relative to other immunoassays and is a convenient and sensitive semiautomated method to measure CsA in whole blood. Copyright © 2010 by Lippincott Williams & Wilkins.
2010
32
2
145
151
Cyclosporine; Immunoassay; LC/MS/MS; Method evaluation; Therapeutic drug monitoring
Wallemacq P.; Maine G.T.; Berg K.; Rosiere T.; Marquet P.; Aimo G.; Mengozzi G.; Young J.; Wonigeit K.; Kretschmer R.; Wermuth B.; Schmid R.W.
File in questo prodotto:
File Dimensione Formato  
Multisite_Architect.pdf

Accesso aperto

Dimensione 231.06 kB
Formato Adobe PDF
231.06 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1890019
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 31
  • ???jsp.display-item.citation.isi??? 24
social impact