Patients undergoing explantation of left ventricular assist devices (LVADs) after improvement of myocardial function remain a minority. Nevertheless, considering the growing population of LVAD patients, increasing demand for new explantation strategies is expected. Herein, we present a retrospective review of seven patients undergoing HeartMate3 explantation with the use of a custom-made apical ring plug in four medical centers. The primary outcome was status at intensive care unit discharge. Secondary outcomes included perioperative complications and transfusions. Six out of seven patients were males. The median age at explantation and time on LVAD support was 35 years (range:13-73) and 10 months (range:9-24), respectively. No technical difficulties were experienced during plug implantation via a conventional sternotomy or through a left lateral thoracotomy, either with or without cardiopulmonary bypass. Perioperative transfusions ranged from 0 to 3 units/patient. No re-operations for bleeding, hemorrhagic, embolic, or plug-related infective events were observed. Heparin was started 6 hours after surgery as a bridge to oral anticoagulation (international normalized ratio: 2-2.5). All patients were discharged alive from intensive care unit. This novel plug device for HeartMate3 explantation was successfully and safely implanted in this first patient series. Notwithstanding, its use should still be considered off-label and larger studies are required to investigate its long-term results.

Novel Plug Device for HeartMate 3 Explantation: First Multicenter Experience.

Loforte A;
2022-01-01

Abstract

Patients undergoing explantation of left ventricular assist devices (LVADs) after improvement of myocardial function remain a minority. Nevertheless, considering the growing population of LVAD patients, increasing demand for new explantation strategies is expected. Herein, we present a retrospective review of seven patients undergoing HeartMate3 explantation with the use of a custom-made apical ring plug in four medical centers. The primary outcome was status at intensive care unit discharge. Secondary outcomes included perioperative complications and transfusions. Six out of seven patients were males. The median age at explantation and time on LVAD support was 35 years (range:13-73) and 10 months (range:9-24), respectively. No technical difficulties were experienced during plug implantation via a conventional sternotomy or through a left lateral thoracotomy, either with or without cardiopulmonary bypass. Perioperative transfusions ranged from 0 to 3 units/patient. No re-operations for bleeding, hemorrhagic, embolic, or plug-related infective events were observed. Heparin was started 6 hours after surgery as a bridge to oral anticoagulation (international normalized ratio: 2-2.5). All patients were discharged alive from intensive care unit. This novel plug device for HeartMate3 explantation was successfully and safely implanted in this first patient series. Notwithstanding, its use should still be considered off-label and larger studies are required to investigate its long-term results.
2022
68(12)
262
267
https://pubmed.ncbi.nlm.nih.gov/35544446/
Mariani S; Li T; Meyer AL; Geoffrion T; Fittkau M; Schöde A; Hanke JS; Loforte A; Haverich A; Schmitto JD; Dogan G.
File in questo prodotto:
File Dimensione Formato  
Novel_Plug_Device_for_HeartMate_3_Explantation_.25.pdf

Accesso riservato

Descrizione: Plug recovery HM3
Tipo di file: PDF EDITORIALE
Dimensione 445.3 kB
Formato Adobe PDF
445.3 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1898332
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? ND
social impact