According to recent investigations, the risk of M. chimaera contamination of heater-cooler units (HCUs) has reached global proportions. Our aim was to field evaluate a protocol for early detection of M. chimaera contamination. We assessed the presence of viable M. chimaera in 395 water samples obtained from 48 devices (HCUs and extracorporeal membrane oxygenation) by Real Time PCR. Thirty devices were NTM positive, of which 14 were contaminated with M. chimaera. The most frequently contaminated devices were the Stockert 3T. Noteworthy, Stockert 3T devices were positive for M. chimaera. In conclusion, this study introduces novel PMA-PCR designed to specifically detect M. chimaera in HCUs and ECMO devices; this method can replace the culture method for continuous microbiological surveillance. The timely detection of M. chimaera contamination can then be used to improve effective management of the devices.

Real-time PCR, the best approaches for rapid testing for Mycobacterium chimaera detection in heater cooler units and extracorporeal membrane oxygenation

Ditommaso S.
;
Giacomuzzi M.;Zotti C. M.
2021-01-01

Abstract

According to recent investigations, the risk of M. chimaera contamination of heater-cooler units (HCUs) has reached global proportions. Our aim was to field evaluate a protocol for early detection of M. chimaera contamination. We assessed the presence of viable M. chimaera in 395 water samples obtained from 48 devices (HCUs and extracorporeal membrane oxygenation) by Real Time PCR. Thirty devices were NTM positive, of which 14 were contaminated with M. chimaera. The most frequently contaminated devices were the Stockert 3T. Noteworthy, Stockert 3T devices were positive for M. chimaera. In conclusion, this study introduces novel PMA-PCR designed to specifically detect M. chimaera in HCUs and ECMO devices; this method can replace the culture method for continuous microbiological surveillance. The timely detection of M. chimaera contamination can then be used to improve effective management of the devices.
2021
36
6
626
629
Mycobacterium chimaera; PCR; cardiac surgery; disinfection; heater-cooler unit; microbiological monitoring
Ditommaso S.; Giacomuzzi M.; Memoli G.; Zotti C.M.
File in questo prodotto:
File Dimensione Formato  
paper ditommaso revised.docx

Accesso riservato

Dimensione 50.16 kB
Formato Microsoft Word XML
50.16 kB Microsoft Word XML   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1933910
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 3
social impact