AimsDespite the high prevalence rate of atrial high-rate episodes (AHREs) detected using cardiac implantable electronic devices (CIEDs), clinical guidelines and consensus documents have disagreed on a universal AHRE definition and a temporal cut-off related to subsequent thromboembolic events. This diagnostic test accuracy meta-analysis aims to derive the optimal temporal threshold of clinically significant AHREs from the available literature.MethodsThe PubMed/MEDLINE and EMBASE databases were screened for studies on CIED patients reporting the incidence of thromboembolic events related to at least one AHRE temporal cut-off. A total of 23 studies were included: 19 considering the longest single AHRE and four the AHRE burden, respectively. A random-effect diagnostic test accuracy meta-analysis with multiple cut-offs was performed. Two analyses were performed according to the AHRE temporal cut-off subtype (longest episode vs. cumulative burden).ResultsThe analysis on the longest single AHRE indicated 0.07 min as the optimal duration to differentiate AHRE associated or not with thromboembolic events [sensitivity 65.4% (95% CI 48.8%-79.0%), specificity 52.7% (95% CI 46.0%-59.4%), and area under the summary receiver operating characteristic curve (AUC-SROC): 0.62]. The analysis on AHRE burden indicated 1.4 min as the optimal cut-off [sensitivity 58.2% (95% CI 25.6%-85.0%), specificity 57.5% (95% CI 42.0%-71.7%), and AUC-SROC 0.60]. A sensitivity analysis excluding patients with a history of atrial fibrillation and including high-quality studies only yielded similar results.ConclusionThe presence of AHRE, rather than a specific duration, relates to an increased, albeit low, thromboembolic risk in CIED patients. Any AHRE should constitute an additional element in patient-specific thromboembolic risk assessment.Graphical abstract reporting the main findings of the diagnostic meta-analysis.

Role of atrial high-rate episodes in stratifying thromboembolic risk: a multiple cut-off diagnostic meta-analysis

Saglietto, Andrea
First
;
Ballatore, Andrea;Griffith Brookles, Carola;Xhakupi, Henri;De Ferrari, Gaetano Maria
;
Anselmino, Matteo
Last
2023-01-01

Abstract

AimsDespite the high prevalence rate of atrial high-rate episodes (AHREs) detected using cardiac implantable electronic devices (CIEDs), clinical guidelines and consensus documents have disagreed on a universal AHRE definition and a temporal cut-off related to subsequent thromboembolic events. This diagnostic test accuracy meta-analysis aims to derive the optimal temporal threshold of clinically significant AHREs from the available literature.MethodsThe PubMed/MEDLINE and EMBASE databases were screened for studies on CIED patients reporting the incidence of thromboembolic events related to at least one AHRE temporal cut-off. A total of 23 studies were included: 19 considering the longest single AHRE and four the AHRE burden, respectively. A random-effect diagnostic test accuracy meta-analysis with multiple cut-offs was performed. Two analyses were performed according to the AHRE temporal cut-off subtype (longest episode vs. cumulative burden).ResultsThe analysis on the longest single AHRE indicated 0.07 min as the optimal duration to differentiate AHRE associated or not with thromboembolic events [sensitivity 65.4% (95% CI 48.8%-79.0%), specificity 52.7% (95% CI 46.0%-59.4%), and area under the summary receiver operating characteristic curve (AUC-SROC): 0.62]. The analysis on AHRE burden indicated 1.4 min as the optimal cut-off [sensitivity 58.2% (95% CI 25.6%-85.0%), specificity 57.5% (95% CI 42.0%-71.7%), and AUC-SROC 0.60]. A sensitivity analysis excluding patients with a history of atrial fibrillation and including high-quality studies only yielded similar results.ConclusionThe presence of AHRE, rather than a specific duration, relates to an increased, albeit low, thromboembolic risk in CIED patients. Any AHRE should constitute an additional element in patient-specific thromboembolic risk assessment.Graphical abstract reporting the main findings of the diagnostic meta-analysis.
2023
10
1289372
1
10
AHRE; atrial fibrillation; cardiac implantable electronic device; subclinical AF; thromboembolic risk
Saglietto, Andrea; Ballatore, Andrea; Griffith Brookles, Carola; Xhakupi, Henri; De Ferrari, Gaetano Maria; Anselmino, Matteo
File in questo prodotto:
File Dimensione Formato  
Saglietto Front Cardiovasc Med 2023.pdf

Accesso aperto

Tipo di file: PDF EDITORIALE
Dimensione 2.35 MB
Formato Adobe PDF
2.35 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1946259
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact