Objective This study aimed to investigate the influence of cup positioning accuracy of two different cup impactors handled by two surgeons with different levels of total hip replacement surgery experience.Study Design This was a cadaveric study ( n = 20). A standard and a new prototype Kyon impactor were used by two surgeons (beginner and senior) to impact 80 prosthetic cups in 40 canine normal acetabula. The angle of lateral opening, angle of cup retroversion, and their corresponding deviations from the ideal postoperative values were measured using radiographic imaging to evaluate the accuracy of cup positioning.Results The accuracy of cup positioning was not affected by the type of impactor used by the senior surgeon. For the beginner surgeon, the standard impactor was associated with a significantly higher mean error (3.9 +/- 6.5 degrees) and mean value (48.90 +/- 6.5 degrees) for the angle of lateral opening compared with the prototype impactor that showed a lower mean error (-1.8 +/- 6.3 degrees) and mean values (43.20 +/- 6.34 degrees) closer to the ideal value of 45 degrees. No significant differences were found between the standard impactor and the prototype regarding the accuracy of the angle of cup retroversion.Conclusion The new features of the prototype impactor and its improved handling can assist inexperienced surgeons in achieving a more accurate angle of lateral opening compared with the standard impactor. In contrast, the accuracy of the cup retroversion angle appears to be more closely related to the surgeon's experience rather than the type of impactor used.
Comparison between Two Zurich Cementless Total Hip Replacement Cup Impactor Types in the Accuracy of Cup Positioning: An In Vitro Study
Mancusi D.
First
;Peirone B.;Terrana M. R.;Piras L. A.Last
2025-01-01
Abstract
Objective This study aimed to investigate the influence of cup positioning accuracy of two different cup impactors handled by two surgeons with different levels of total hip replacement surgery experience.Study Design This was a cadaveric study ( n = 20). A standard and a new prototype Kyon impactor were used by two surgeons (beginner and senior) to impact 80 prosthetic cups in 40 canine normal acetabula. The angle of lateral opening, angle of cup retroversion, and their corresponding deviations from the ideal postoperative values were measured using radiographic imaging to evaluate the accuracy of cup positioning.Results The accuracy of cup positioning was not affected by the type of impactor used by the senior surgeon. For the beginner surgeon, the standard impactor was associated with a significantly higher mean error (3.9 +/- 6.5 degrees) and mean value (48.90 +/- 6.5 degrees) for the angle of lateral opening compared with the prototype impactor that showed a lower mean error (-1.8 +/- 6.3 degrees) and mean values (43.20 +/- 6.34 degrees) closer to the ideal value of 45 degrees. No significant differences were found between the standard impactor and the prototype regarding the accuracy of the angle of cup retroversion.Conclusion The new features of the prototype impactor and its improved handling can assist inexperienced surgeons in achieving a more accurate angle of lateral opening compared with the standard impactor. In contrast, the accuracy of the cup retroversion angle appears to be more closely related to the surgeon's experience rather than the type of impactor used.| File | Dimensione | Formato | |
|---|---|---|---|
|
b)Main document.docx
Accesso aperto
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
78.15 kB
Formato
Microsoft Word XML
|
78.15 kB | Microsoft Word XML | Visualizza/Apri |
|
Abstract.docx
Accesso aperto
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
15.14 kB
Formato
Microsoft Word XML
|
15.14 kB | Microsoft Word XML | Visualizza/Apri |
|
a) Title Page.docx
Accesso aperto
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
16.96 kB
Formato
Microsoft Word XML
|
16.96 kB | Microsoft Word XML | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.