The aims of this study were: (1) to develop and assess reproducibility of a new method for measuring masticatory force in the intercuspal position; (2) to test the reproducibility of surface EMG signal amplitude and spectral variables in constant force contractions of jaw elevator muscles and its dependency on inter-electrode distance. The study was performed on the masseter and temporalis anterior muscles of both sides of nine healthy volunteers. An intraoral compressive-force sensor was used to measure maximal voluntary contraction forces in the intercuspal position and to provide a visual feedback on sub-maximal forces to the subject. Three experimental sessions were performed in three days. In each session, three isometric contractions at 80% of the maximal force were sustained by the subjects for 30s. The intra-class correlation coefficient (ICC) of the maximal force measure was 71.9%. ICC of average rectified value and mean power spectral frequency of the EMG signal increased with inter-electrode distance, with values larger than 70% with 30 mm inter-electrode distance. It was concluded that surface EMG variables measured in isometric contractions of the jaw elevator muscles with the proposed force recording system show good reproducibility for clinical applications when a 30 mm inter-electrode distance is considered.
Reproducibility of surface EMG variables in isometric sub-maximal contractions of jaw elevator muscles
CASTROFLORIO, Tommaso;DEBERNARDI, Cesare Lorenzo;BRACCO, Pietro;
2006-01-01
Abstract
The aims of this study were: (1) to develop and assess reproducibility of a new method for measuring masticatory force in the intercuspal position; (2) to test the reproducibility of surface EMG signal amplitude and spectral variables in constant force contractions of jaw elevator muscles and its dependency on inter-electrode distance. The study was performed on the masseter and temporalis anterior muscles of both sides of nine healthy volunteers. An intraoral compressive-force sensor was used to measure maximal voluntary contraction forces in the intercuspal position and to provide a visual feedback on sub-maximal forces to the subject. Three experimental sessions were performed in three days. In each session, three isometric contractions at 80% of the maximal force were sustained by the subjects for 30s. The intra-class correlation coefficient (ICC) of the maximal force measure was 71.9%. ICC of average rectified value and mean power spectral frequency of the EMG signal increased with inter-electrode distance, with values larger than 70% with 30 mm inter-electrode distance. It was concluded that surface EMG variables measured in isometric contractions of the jaw elevator muscles with the proposed force recording system show good reproducibility for clinical applications when a 30 mm inter-electrode distance is considered.File | Dimensione | Formato | |
---|---|---|---|
JEK_2006.pdf
Accesso riservato
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
237.36 kB
Formato
Adobe PDF
|
237.36 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.