Dose distribution in mantle fields irradiated with 18 MeV photons has been studied in an Alderson-Rando phantom with thermoluminescent dosimeters. The treatment was also simulated on an RT-plan 7000 computer; calculated and measured doses in anatomical sites of major interest were compared. Doses in all significant points were within +/- 7% of the doses at the reference point. Calculated and measured doses agreed within +/- 5% except points situated near unhomogeneous tissues. The authors discuss their results and present conclusions about algorithms for computer dose calculation in irregular field treatments.
[Mantle irradiation with 18 MeV photons. Measurement and calculation of dose distribution]
RAGONA, Riccardo;
1985-01-01
Abstract
Dose distribution in mantle fields irradiated with 18 MeV photons has been studied in an Alderson-Rando phantom with thermoluminescent dosimeters. The treatment was also simulated on an RT-plan 7000 computer; calculated and measured doses in anatomical sites of major interest were compared. Doses in all significant points were within +/- 7% of the doses at the reference point. Calculated and measured doses agreed within +/- 5% except points situated near unhomogeneous tissues. The authors discuss their results and present conclusions about algorithms for computer dose calculation in irregular field treatments.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.