We evaluated the influence of manual preflaring and torque on the failure rate of rotary nickel-titanium ProTaper instruments Shaping 1 (S1), Shaping 2 (S2), Finishing 1 (F1), and Finishing 2 (F2). These factors were evaluated using an in vitro method by calculating the mean number of Endo-Training-Blocks shaped before file breakage under different conditions. Group A (S1 on simulators with no preflaring) shaped 10 blocks before failure, group B (S1 on manually preflared simulators) shaped 59 blocks (p<0.01 versus group A), group C (S2 with low torque) shaped 28 blocks, group D (S2 with high torque) shaped 48 blocks (p<0.01 versus group C), group E (F1 with low torque) shaped eight blocks, group F (F1 with high torque) shaped 23 blocks (p<0.01 versus group E), group G (F2 with low torque) shaped four blocks, and group H (F2 with high torque) shaped 11 blocks (p<0.01 versus group G). Manual preflaring creates a glide path for the instrument tip and is a major determinant in reducing the failure rate of these rotary nickel-titanium files. All instruments worked better at high torque.

Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments

BERUTTI, Elio;PASQUALINI, Damiano
2004-01-01

Abstract

We evaluated the influence of manual preflaring and torque on the failure rate of rotary nickel-titanium ProTaper instruments Shaping 1 (S1), Shaping 2 (S2), Finishing 1 (F1), and Finishing 2 (F2). These factors were evaluated using an in vitro method by calculating the mean number of Endo-Training-Blocks shaped before file breakage under different conditions. Group A (S1 on simulators with no preflaring) shaped 10 blocks before failure, group B (S1 on manually preflared simulators) shaped 59 blocks (p<0.01 versus group A), group C (S2 with low torque) shaped 28 blocks, group D (S2 with high torque) shaped 48 blocks (p<0.01 versus group C), group E (F1 with low torque) shaped eight blocks, group F (F1 with high torque) shaped 23 blocks (p<0.01 versus group E), group G (F2 with low torque) shaped four blocks, and group H (F2 with high torque) shaped 11 blocks (p<0.01 versus group G). Manual preflaring creates a glide path for the instrument tip and is a major determinant in reducing the failure rate of these rotary nickel-titanium files. All instruments worked better at high torque.
2004
30
228
230
http://journals.elsevierhealth.com/periodicals/ymai/medline/record/MDLN.15085052
Dental Alloys; Dental Instruments; Dental Models; Equipment Failure; Nickel; Root Canal Preparation; Titanium; Torque
E. BERUTTI; A.R. NEGRO; M. LENDINI; D. PASQUALINI
File in questo prodotto:
File Dimensione Formato  
Berutti_JE2004.pdf

Accesso riservato

Tipo di file: POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione 52.17 kB
Formato Adobe PDF
52.17 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/38579
Citazioni
  • ???jsp.display-item.citation.pmc??? 35
  • Scopus 140
  • ???jsp.display-item.citation.isi??? 116
social impact