This work aimed at identifying the main sources of uncertainty for the measurement of dissolved oxygen concentration in aqueous solutions. The experimental apparatus consists of an amperometric cell based on the Clark-type sensor. The corresponding uncertainty budget was assessed, this being a fundamental step for the validation of a measurement method. The principle of the measurement, as well as the procedure for the set-up and the characterisation of the cell, are described. The measurement equation was defined as a combination of Faraday’s and Fick’s laws, and a method was worked out for the empirical determination of the diffusivity parameter. In this connection, the solutions of oxygen were standardised by way of the Winkler’s titration, as suggested by the ISO Guide 5813 and 5814. With this approach we aimed at contributing to the development of a potential primary method of measurement. A discussion of all the contributions to the overall uncertainty is reported, allowing operators to locate the largest ones and plan specific improvements.

Assessment of the uncertainty budget for the amperometric measurement of dissolved oxygen

PRENESTI, Enrico
2007-01-01

Abstract

This work aimed at identifying the main sources of uncertainty for the measurement of dissolved oxygen concentration in aqueous solutions. The experimental apparatus consists of an amperometric cell based on the Clark-type sensor. The corresponding uncertainty budget was assessed, this being a fundamental step for the validation of a measurement method. The principle of the measurement, as well as the procedure for the set-up and the characterisation of the cell, are described. The measurement equation was defined as a combination of Faraday’s and Fick’s laws, and a method was worked out for the empirical determination of the diffusivity parameter. In this connection, the solutions of oxygen were standardised by way of the Winkler’s titration, as suggested by the ISO Guide 5813 and 5814. With this approach we aimed at contributing to the development of a potential primary method of measurement. A discussion of all the contributions to the overall uncertainty is reported, allowing operators to locate the largest ones and plan specific improvements.
2007
597
75
81
Clark-type sensors; Amperometric measurement; Dissolved oxygen measurement uncertainty; validation
P. FISICARO; A. ADRIAENS; E. FERRARA; E. PRENESTI
File in questo prodotto:
File Dimensione Formato  
oxyegn dissolved.pdf

Accesso aperto

Tipo di file: POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione 392.92 kB
Formato Adobe PDF
392.92 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/99557
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 16
social impact