Background Finding of intracardiac lead masses in patients with cardiac implantable electronic devices remains controversial, as such masses have been observed in cases of exclusively local infections whereas they have not been recognized in patients with positive cultures of intravascular lead fragments. In this study, we aim to describe the prevalence of intracardiac lead masses in true asymptomatic patients with cardiac implantable electronic devices, to identify their predictive factors and to define their prognostic impact at long-term follow-up. Methods Seventy-eight consecutive patients admitted over a 6-month period for elective generator replacement without clinical evidence of infection were evaluated by transthoracic and transesophageal echocardiography and prospectively followed at in-clinic follow-up visits. Results Lead masses were found in 10 patients (12.8%). These patients had more frequently right ventricular dysfunction at univariate analysis (OR 2.71, PU0.010) and after baseline variables adjustment (hazard ratio 6.25, PU0.012). At 5-year follow-up without any specific therapy, none of the patients suffered from any cardiac device infections, or developed clinical signs of infections. Conclusion There is an evidence of clinical lead masses in asymptomatic patients with cardiac implantable electronic devices. The value of these findings is still debated for aetiological interpretation and for therapeutic strategy, but they are not necessarily associated with an infection.
Prevalence and prognosis of lead masses in patients with cardiac implantable electronic devices without infection
Errigo, Daniele;Peyracchia, Mattia;Frea, Simone;Castagno, Davide;Budano, Carlo;Giustetto, Carla;Rinaldi, Mauro
Last
2019-01-01
Abstract
Background Finding of intracardiac lead masses in patients with cardiac implantable electronic devices remains controversial, as such masses have been observed in cases of exclusively local infections whereas they have not been recognized in patients with positive cultures of intravascular lead fragments. In this study, we aim to describe the prevalence of intracardiac lead masses in true asymptomatic patients with cardiac implantable electronic devices, to identify their predictive factors and to define their prognostic impact at long-term follow-up. Methods Seventy-eight consecutive patients admitted over a 6-month period for elective generator replacement without clinical evidence of infection were evaluated by transthoracic and transesophageal echocardiography and prospectively followed at in-clinic follow-up visits. Results Lead masses were found in 10 patients (12.8%). These patients had more frequently right ventricular dysfunction at univariate analysis (OR 2.71, PU0.010) and after baseline variables adjustment (hazard ratio 6.25, PU0.012). At 5-year follow-up without any specific therapy, none of the patients suffered from any cardiac device infections, or developed clinical signs of infections. Conclusion There is an evidence of clinical lead masses in asymptomatic patients with cardiac implantable electronic devices. The value of these findings is still debated for aetiological interpretation and for therapeutic strategy, but they are not necessarily associated with an infection.File | Dimensione | Formato | |
---|---|---|---|
Golzio et al-2019 JCM-Prevalence and prognosis of lead masses.docx
Accesso riservato
Descrizione: Articolo completo
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
1.04 MB
Formato
Microsoft Word XML
|
1.04 MB | Microsoft Word XML | Visualizza/Apri Richiedi una copia |
Golzio et al-2019 JCM-Prevalence and prognosis of lead masses.pdf
Open Access dal 02/06/2020
Descrizione: Articolo completo
Tipo di file:
POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione
615.4 kB
Formato
Adobe PDF
|
615.4 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.