Sensor technology was introduced to intraoperatively analyse the differential pressure between the medial and lateral compartments of the knee during primary TKA using a sensor to assess if further balancing procedures are needed to achieve a “balanced” knee. The prognostic role of epidemiological and radiological parameters was also analysed. A consecutive series of 21 patients with primary knee osteoarthritis were enrolled and programmed for TKA in our unit between 1 September 2020 and 31 March 2021. The VERASENSE Knee System (OrthoSensor Inc., Dania Beach, FL, USA) has been proposed as an instrument that quantifies the differential pressure between the compartments of the knee intraoperatively throughout the full range of motion during primary TKA, designed with a J‐curve anatomical femoral design and a PS “medially congruent” polyethylene insert. Thirteen patients (61.90%) showed a “balanced” knee, and eight patients (38.10%) showed an intra‐operative “unbalanced” knee and required additional procedures. A total of 13 additional balancing procedures were performed. At the end of surgical knee procedures, a quantitatively balanced knee was obtained in all patients. In addition, a correlation was found between the compartment pressure of phase I and phase II at 10° of flexion and higher absolute pressures were found in the medial compartment than in the lateral compartment in each ROM degree investigated. Moreover, those pressure values showed a trend to decrease with the increase in flexion degrees in both compartments. The “Kinetic Tracking” function displays the knee’s dynamic motion through the full ROM to evaluate joint kinetics. The obtained kinetic traces reproduced the knee’s medial pivot and femoral rollback, mimicking natural knee biomechanics. Moreover, we re-ported a statistically significant correlation between the need for soft tissue or bone resection re-balancing and severity of the initial coronal deformity (>10°) and a preoperative JLCA value >2°. The use of quantitative sensor‐guided pressure evaluation during TKA leads to a more reproducible “balanced” knee. The surgeon, evaluating radiological parameters before surgery, may anticipate difficulties in knee balance and require those devices to achieve the desired result objectively.

Kinetic sensors for ligament balance and kinematic evaluation in anatomic bi‐cruciate stabilized total knee arthroplasty

Sabatini L.;Bosco F.;Barberis L.;Camazzola D.;Bistolfi A.;Masse A.;
2021-01-01

Abstract

Sensor technology was introduced to intraoperatively analyse the differential pressure between the medial and lateral compartments of the knee during primary TKA using a sensor to assess if further balancing procedures are needed to achieve a “balanced” knee. The prognostic role of epidemiological and radiological parameters was also analysed. A consecutive series of 21 patients with primary knee osteoarthritis were enrolled and programmed for TKA in our unit between 1 September 2020 and 31 March 2021. The VERASENSE Knee System (OrthoSensor Inc., Dania Beach, FL, USA) has been proposed as an instrument that quantifies the differential pressure between the compartments of the knee intraoperatively throughout the full range of motion during primary TKA, designed with a J‐curve anatomical femoral design and a PS “medially congruent” polyethylene insert. Thirteen patients (61.90%) showed a “balanced” knee, and eight patients (38.10%) showed an intra‐operative “unbalanced” knee and required additional procedures. A total of 13 additional balancing procedures were performed. At the end of surgical knee procedures, a quantitatively balanced knee was obtained in all patients. In addition, a correlation was found between the compartment pressure of phase I and phase II at 10° of flexion and higher absolute pressures were found in the medial compartment than in the lateral compartment in each ROM degree investigated. Moreover, those pressure values showed a trend to decrease with the increase in flexion degrees in both compartments. The “Kinetic Tracking” function displays the knee’s dynamic motion through the full ROM to evaluate joint kinetics. The obtained kinetic traces reproduced the knee’s medial pivot and femoral rollback, mimicking natural knee biomechanics. Moreover, we re-ported a statistically significant correlation between the need for soft tissue or bone resection re-balancing and severity of the initial coronal deformity (>10°) and a preoperative JLCA value >2°. The use of quantitative sensor‐guided pressure evaluation during TKA leads to a more reproducible “balanced” knee. The surgeon, evaluating radiological parameters before surgery, may anticipate difficulties in knee balance and require those devices to achieve the desired result objectively.
2021
Inglese
Esperti anonimi
21
16
1
13
13
Article nr. 5427
Femoral rollback; Kinetic sensor; Knee balance; Medial pivot; Predictor variables; Total knee arthroplasty; Biomechanical Phenomena; Humans; Kinetics; Knee Joint; Ligaments; Range of Motion, Articular; Arthroplasty, Replacement, Knee; Knee Prosthesis; Osteoarthritis, Knee
no
1 – prodotto con file in versione Open Access (allegherò il file al passo 6 - Carica)
262
8
Sabatini L.; Bosco F.; Barberis L.; Camazzola D.; Bistolfi A.; Risitano S.; Masse A.; Indelli P.F.
info:eu-repo/semantics/article
open
03-CONTRIBUTO IN RIVISTA::03A-Articolo su Rivista
File in questo prodotto:
File Dimensione Formato  
sensors-21-05427-v2.pdf

Accesso aperto

Tipo di file: PDF EDITORIALE
Dimensione 1.02 MB
Formato Adobe PDF
1.02 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1841060
Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 14
social impact