Objective: We introduce a new treatment verification technique to estimate the primary particle's stopping power from prompt gamma timing measurements in proton therapy. Approach: The starting point is the Spatio-temporal Emission Recostruction technique, which provides the time-depth distribution of the emitted prompt photons with a multiple Prompt-Gamma Timing detector setup based on Lanthanum Bromide crystals. A dedicated formalism based on an analytical approximation of the stopping power is developed to obtain the desired information. Its performance is evaluated in a proof of concept configuration via Monte Carlo simulations of monochromatic proton beams impinging on a homogeneous PMMA phantom. Main Results: Results indicate stopping power estimations as good as 3.8% with respect to NIST values, and range estimations within 0.3 cm (standard deviation), when considering 250 ps FWHM timing resolution. Significance: The current study shows, for the first time, the feasibility of evaluating the stopping power of primary beams with a technique that can be performed in-vivo, opening up new possibilities in the field of treatment verification and therapy optimization.

Estimating the stopping power distribution during proton therapy: A proof of concept

Veronica Ferrero
First
;
Piergiorgio Cerello;Elisa Fiorina;Anna Vignati;Francesco Pennazio;Magdalena Rafecas
2022-01-01

Abstract

Objective: We introduce a new treatment verification technique to estimate the primary particle's stopping power from prompt gamma timing measurements in proton therapy. Approach: The starting point is the Spatio-temporal Emission Recostruction technique, which provides the time-depth distribution of the emitted prompt photons with a multiple Prompt-Gamma Timing detector setup based on Lanthanum Bromide crystals. A dedicated formalism based on an analytical approximation of the stopping power is developed to obtain the desired information. Its performance is evaluated in a proof of concept configuration via Monte Carlo simulations of monochromatic proton beams impinging on a homogeneous PMMA phantom. Main Results: Results indicate stopping power estimations as good as 3.8% with respect to NIST values, and range estimations within 0.3 cm (standard deviation), when considering 250 ps FWHM timing resolution. Significance: The current study shows, for the first time, the feasibility of evaluating the stopping power of primary beams with a technique that can be performed in-vivo, opening up new possibilities in the field of treatment verification and therapy optimization.
2022
10
01
09
https://doi.org/10.3389/fphy.2022.971767
treatment verification; prompt gamma timing; reconstruction algorithms; stoppingpower; adaptive treatment; Monte Carlo simulations; proton therapy; rangemonitoring
Veronica Ferrero; Julius Werner; Piergiorgio Cerello; Elisa Fiorina; Anna Vignati; Francesco Pennazio; Magdalena Rafecas
File in questo prodotto:
File Dimensione Formato  
2022Frontiers_MERLINO.pdf

Accesso aperto

Dimensione 1.5 MB
Formato Adobe PDF
1.5 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/1878343
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 3
social impact