Introduction: Nasal synchronized intermittent positive pressure ventilation (nSIPPV) is an effective non-invasive ventilation technique, especially for preterm infants. Patient-ventilator synchrony is essential for providing effective respiratory support; however, no automated system is currently available for monitoring this parameter. A new tool for automatic assessment of patient-ventilator synchrony, the SyncNIV system, was developed and applied in this pilot study to evaluate differences between nSIPPV and non-synchronized nasal intermittent positive pressure ventilation (nIPPV) in preterm infants with respiratory distress. Methods: This study involved designing a custom algorithm for signal analysis. Data were collected through a polygraph that could simultaneously gather respiratory data from the patients and the ventilator. Patient-ventilator synchrony was evaluated by applying the SyncNIV system in a randomized crossover study designed to compare nSIPPV and nIPPV. The primary outcome was the mean instant Synchrony Index (i-SI), defined as the portion of the inspiration effort sustained by ventilator inflation, expressed as a percentage. Results: Fourteen infants with a median (IQR) gestational age of 28.6 (25.6-30.3) were enrolled. We analyzed 43,304 ventilator inflations and 50,221 patient breaths. The i-SI was 54.69% (44.49-60.09) in nSIPPV and 39.54% (33.40-48.75) in nIPPV, p < 0.05. Conclusion: The SyncNIV system confirmed better i-SI during nSIPPV than during nIPPV, demonstrating its effectiveness in assessing the differences between these two modes of non-invasive ventilation in preterm infants. The SyncNIV system could be a useful tool for optimizing the ventilation parameters and improving the effectiveness and comfort of respiratory support systems.
A New Tool to Assess Patient-Ventilator Synchrony in Preterm Infants Receiving Non-Invasive Ventilation: A Randomized Crossover Pilot Study
Cresi, Francesco;Maggiora, Elena;Rubino, Carlotta;Ferroglio, Mattia;Ruzzante, Elena;Giraudo, Isaac;Limone, Marco;Coscia, Alessandra
2025-01-01
Abstract
Introduction: Nasal synchronized intermittent positive pressure ventilation (nSIPPV) is an effective non-invasive ventilation technique, especially for preterm infants. Patient-ventilator synchrony is essential for providing effective respiratory support; however, no automated system is currently available for monitoring this parameter. A new tool for automatic assessment of patient-ventilator synchrony, the SyncNIV system, was developed and applied in this pilot study to evaluate differences between nSIPPV and non-synchronized nasal intermittent positive pressure ventilation (nIPPV) in preterm infants with respiratory distress. Methods: This study involved designing a custom algorithm for signal analysis. Data were collected through a polygraph that could simultaneously gather respiratory data from the patients and the ventilator. Patient-ventilator synchrony was evaluated by applying the SyncNIV system in a randomized crossover study designed to compare nSIPPV and nIPPV. The primary outcome was the mean instant Synchrony Index (i-SI), defined as the portion of the inspiration effort sustained by ventilator inflation, expressed as a percentage. Results: Fourteen infants with a median (IQR) gestational age of 28.6 (25.6-30.3) were enrolled. We analyzed 43,304 ventilator inflations and 50,221 patient breaths. The i-SI was 54.69% (44.49-60.09) in nSIPPV and 39.54% (33.40-48.75) in nIPPV, p < 0.05. Conclusion: The SyncNIV system confirmed better i-SI during nSIPPV than during nIPPV, demonstrating its effectiveness in assessing the differences between these two modes of non-invasive ventilation in preterm infants. The SyncNIV system could be a useful tool for optimizing the ventilation parameters and improving the effectiveness and comfort of respiratory support systems.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.