Comprehensive two-dimensional gas chromatography (GC×GC) coupled with Mass Spectrometry (MS) is one of today’s most powerful analytical platforms for detailed analysis of medium-to-high complexity samples. The column set usually consists of a long, conventional-inner-diameter first dimension (1D) (typically 15-30 m long, 0.32-0.25 mm dc), and a short, narrow-bore second dimension (2D) column (typically 0.5-2 m, 0.1 mm dc) where separation is run in a few seconds. However, when thermal modulation is used, since the columns of a set are coupled in series, a flow mismatch occurs between the two dimensions, making it impossible to operate simultaneously at optimized flow conditions. Further, short narrow-bore capillaries can easily be overloaded, because of their lower loadability, limiting the effectiveness of 2D separation. In this study, improved gas linear velocities in both chromatographic dimensions were achieved by coupling the 1D column with two parallel 2D columns, having identical inner diameter, stationary phase chemistry, and film thickness. In turn, these were connected to two detectors: a fast quadrupole Mass Spectrometer (MS) and a Flame Ionization Detector (FID). Different configurations were tested and performances compared to a conventional set-up; experimental results on two model mixtures (n-alkanes and fourteen medium-to-high polarity volatiles of interest in the flavor and fragrance field) and on the essential oil of Artemisia umbelliformis Lam., show the system provides consistent results, in terms of analyte identification (reliability of spectra and MS matching) and quantitation, also affording an internal cross-validation of quantitation accuracy.

Parallel dual secondary column-dual detection: A further way of enhancing the informative potential of two-dimensional comprehensive gas chromatography

NICOLOTTI, LUCA;CORDERO, Chiara Emilia Irma;BRESSANELLO, DAVIDE;CAGLIERO, Cecilia Lucia;LIBERTO, Erica;MAGAGNA, FEDERICO;RUBIOLO, Patrizia;SGORBINI, Barbara;BICCHI, Carlo
2014-01-01

Abstract

Comprehensive two-dimensional gas chromatography (GC×GC) coupled with Mass Spectrometry (MS) is one of today’s most powerful analytical platforms for detailed analysis of medium-to-high complexity samples. The column set usually consists of a long, conventional-inner-diameter first dimension (1D) (typically 15-30 m long, 0.32-0.25 mm dc), and a short, narrow-bore second dimension (2D) column (typically 0.5-2 m, 0.1 mm dc) where separation is run in a few seconds. However, when thermal modulation is used, since the columns of a set are coupled in series, a flow mismatch occurs between the two dimensions, making it impossible to operate simultaneously at optimized flow conditions. Further, short narrow-bore capillaries can easily be overloaded, because of their lower loadability, limiting the effectiveness of 2D separation. In this study, improved gas linear velocities in both chromatographic dimensions were achieved by coupling the 1D column with two parallel 2D columns, having identical inner diameter, stationary phase chemistry, and film thickness. In turn, these were connected to two detectors: a fast quadrupole Mass Spectrometer (MS) and a Flame Ionization Detector (FID). Different configurations were tested and performances compared to a conventional set-up; experimental results on two model mixtures (n-alkanes and fourteen medium-to-high polarity volatiles of interest in the flavor and fragrance field) and on the essential oil of Artemisia umbelliformis Lam., show the system provides consistent results, in terms of analyte identification (reliability of spectra and MS matching) and quantitation, also affording an internal cross-validation of quantitation accuracy.
2014
1360
264
274
Two-dimensional comprehensive gaschromatography-mass spectrometry; arallel dual secondary column-dualdetection; Essential oil analysis; Outlet pressure correction; Dual 2D pattern alignment; Second dimension linear velocity optimization
Luca Nicolotti; Chiara Cordero*; Davide Bressanello; Cecilia Cagliero; Erica Liberto;Federico Magagna; Patrizia Rubiolo; Barbara Sgorbini; Carlo Bicchi
File in questo prodotto:
File Dimensione Formato  
OA_parallel dual secondary column dual detection.pdf

Open Access dal 20/09/2016

Tipo di file: POSTPRINT (VERSIONE FINALE DELL’AUTORE)
Dimensione 1.75 MB
Formato Adobe PDF
1.75 MB Adobe PDF Visualizza/Apri
Full text.pdf

Accesso riservato

Tipo di file: PDF EDITORIALE
Dimensione 2.21 MB
Formato Adobe PDF
2.21 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2318/148168
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 30
  • ???jsp.display-item.citation.isi??? 27
social impact